Skip to content

Public FDA record

Z-2038-2024: EsoFLIP, 30mm BALLOON DILATION CATHETER, REF ES-330, Rx Only

Recall record Z-2038-2024 for Covidien, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2038-2024
Company in FDA record
Covidien, LLC
FDA classification date
Jun 7, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements that may lead to various harms.

Source facts

Action
On April 25, 2024, Medtronic issued a "Urgent: Medical Device Recall" Notification. Medtronic ask consignees to take the following actions: 1. Immediately identify and quarantine all unused EsoflipTM dilation catheter 30mm Model # ES-330 from the affected lot numbers. 2. Return all unused affected product(s) in your inventory to: Medtronic Attn: Field Returns Dept. 195 McDermott Road North Haven, CT 06473 USA 3. Credit for returned catheters that are part of this recall will be applied upon receipt. 4. Please complete and return the enclosed Customer Confirmation Form, even if you have no product to return. 5. Product purchased directly from Medtronic, please contact 1-800-448-3644 (option 2 followed by option 2 again) for Return Goods Authorization (RGA). Your local Medtronic Representative can assist you as necessary in initiating the return of this product. 6. If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. 7. This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. 8. Please transfer this notice to other organizations on which this action has an impact. 9. Please maintain a copy of this notice in your records.
Initiated
2024-04-25
Root cause
Nonconforming Material/Component
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Canada, Denmark, Germany, Ireland, Italy, Kuwait, New Zealand, Norway, Portugal, Qatar, South Africa, Switzerland, United Arab Emirates.
Recall number
Z-2038-2024
Recall status
Open, Classified
CDRH record ID
207541
FDA classified
2024-06-07
Recall event number
94541

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗