Open original FDA record ↗Recall record
- FDA identifier
- Z-2039-2009
- Company in FDA record
- Advanced Bionics LLC
- FDA classification date
- Sep 11, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated after Advanced Bionics became aware of fourteen (14) devices manufactured between May 27, 2005 and June 30, 2005 that are confirmed to have or potentially have an issue related to the internal magnet. These devices are were made with an internal magnet where the metal case surrounding the magnet was not properly welded. Patients with these devices may experience a decreas
Source facts
- Action
- Advanced Bionics, LLC issued an "Important Notification" dated June 10, 2009 via Fed Ex to Consignees in the US and Canada. Outside of the US, the notification letters will be distributed in such a manner that delivery confirmation can be obtained. There were three type of notification letters distributed: one for Patients, one for Surgeons and one for Clinicians. Letters described the affected device prompting users on what to do next. An acknowledgement form was enclosed and should be completed and sent via fax to 1-661-362-7621. For further information, contact Advanced Bionics at 1-877-829-0026.
- Initiated
- 2009-06-10
- Root cause
- Process control
- Terminated
- 2012-03-26
- Distribution
- Worldwide Distribution -- US, Canada, France and South Korea.
- Recall number
- Z-2039-2009
- Recall status
- Terminated
- CDRH record ID
- 83675
- FDA classified
- 2009-09-11
- Recall event number
- 52700