Open original FDA record ↗Recall record
- FDA identifier
- Z-2039-2012
- Company in FDA record
- Alere San Diego, Inc.
- FDA classification date
- Jul 19, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
A recall was initiated because Alere San Diego has confirmed that the Triage TOX Drug Screen may have significantly decreased precision and accuracy than claimed. An increased frequency of false positive and/or false negative results has been observed for Triage TOX Drug Screen.
Source facts
- Action
- Alere sent an Urgent Medical Device Recall letter dated June 11, 2012 to all affected customers. The letter informed the customers of the affected product, problem identified and actions to be taken. Customers were instructed to discontinue all use of the recalled lots, immediately inform clinicians of this recall at the testing site(s) and complete and fax the enclosed customer verification form and list of affected lots within 10 days to confirm receipt of the letter. Customers with questions about the information contained in the recall letter, can contact Alere San Diego at 9975 Summers Ridge Road, San Diego, CA 92121 or call (877) 308-8287.
- Initiated
- 2012-05-14
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-12-04
- Distribution
- US Nationwide Distribution
- Recall number
- Z-2039-2012
- Recall status
- Terminated
- CDRH record ID
- 110155
- FDA classified
- 2012-07-19
- Recall event number
- 62293