Open original FDA record ↗Recall record
- FDA identifier
- Z-2039-2014
- Company in FDA record
- Karl Storz Endoscopy
- FDA classification date
- Jul 10, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
KARL STORZ Endoscopy-America is recalling the Essential NEO because the connected power cable could potentially become overheated and result in smoke being generated and the cessation of function of the peripheral that the cable is connected to.
Source facts
- Action
- Storz sent an Safety Alert letter dated June 25, 2014, to all customers who purchased the Essential NEO. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete the attached Acknowledgement/Response Form with contact information and return to KARL STORZ. Customers with questions are instructed to contact KARL STORZ' Technical Support Department at (800) 421-0837, option 3, then option 4.
- Initiated
- 2014-06-25
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-01-13
- Distribution
- Nationwide Distribution including the states of: AL, AZ, CA, CO, CT, District of Columbia, DE, FL, GA, HI, IA, IL, IN, KY, LA, MX, MS, MD, MI, MA, MO, MN. NM, NV, NJ, NY, NC, NE, OK, OH, PA, SC, SD, TX, UT, VA, WI and WA., and Internationally to Canada.
- Recall number
- Z-2039-2014
- Recall status
- Terminated
- CDRH record ID
- 128395
- FDA classified
- 2014-07-10
- Recall event number
- 68646