Z-2040-2021: FilmArray Pneumonia Panel plus (Pneumoplus) Part Numbers: Part No: RFIT-ASY-0143 / UDI: 00815381020314 (

Recall record Z-2040-2021 for BioFire Diagnostics, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2040-2021
Company in FDA record
BioFire Diagnostics, LLC
FDA record creation date
Jul 8, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame.

Source facts

Action
On 06/03/2021, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter via overnight FedEx informing its customer that it has identified a potential for obtaining elevated rates of negative adenovirus results when using the pneumonia panel. The potential false negative results are due to a 10-100x reduction in sensitivity specifically for adenovirus species C when using the pneumonia panel kits within 6 months of expiration. In addition: -The test performance is NOT impacted if kits are more than 6 months from expiration date. -Performance for other adenovirus species (e.g. A, B, D, E, F, and G) are NOT affected -The adenovirus assays on all other of the Recalling Firm's Respiratory Panels are NOT affected. Customer are instructed to: 1) Immediately examine their inventory for product identified in this recall. -If the customer identifies any affected Pneumonia Panels in their inventory (i.e. Pneumonia Panels within 6 months of shelf-life), they may continue using the affected product; however, when adenovirus C infection is suspected, negative results for Adenovirus should be confirmed by another method prior to reporting test results to clinicians, or alternatively, the Recalling Firm's Pneumonia Panel Adenovirus negative results should not be reported. 2) If a customer has further distributed this product, they are to identify their customers and notify them at once of this product recall. 3) Complete the enclosed Acknowledgement of Receipt Form as soon as possible and return it to the Recalling Firm via one of the methods listed on the Acknowledgement of Receipt Form - 1) Fill out electronically using the following URL: (https://www.biofiredx.com/pn-5183/) 2) Scan to PDF and return completed form to: recall5183@biofiredx.com Any questions or assistance - contact customer support department at Support@biofiredx.com or via telephone 1-800-735-6544 option 5.
Initiated
2021-06-07
Root cause
Under Investigation by firm
Terminated
2023-07-31
Distribution
U.S.: AK, AL, AZ, CA, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, IN, KS, KY, KY, LA, MA, MD, ME, MI, MI, MN, MO, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, WI, WV and WY. O.U.S.: Algeria, Angola, Anguilla, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bermuda, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, France, Georgia, Germany, Greece , Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Nepal, Netherlands, Nicaragua, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, United Arab Emirates, and Vietnam.
Recall number
Z-2040-2021
Recall status
Terminated
CDRH record ID
187928
FDA record created
2021-07-08
Recall event number
88117

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗