Skip to content

Public FDA record

Z-2040-2024: MyoSPECT System, Model H3912AA and H3912B, and MyoSPECT ES System, Model H3912AB and H3912BD, dedicated

Recall record Z-2040-2024 for GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2040-2024
Company in FDA record
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
FDA classification date
Jun 7, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential issue regarding the service handles for the internal lead (Pb) covers in that the handles could loosen or release which could cause a cover to drop potentially resulting in injury to service personnel due to the weight of the cover.

Source facts

Action
The recalling firm issued letters on and dated 4/15/2024 via FedEx informing the consignee of the safety issue and actions to be taken by the customer/user. The actions state the consignee can continue to use the system but until GE HealthCare implements the correction for this potential issue on the affected device and before conducting any servicing activities that involve movement of the covers, GE HealthCare Service should be contacted for guidance. An acknowledgement form was enclosed for completion and return. The letter contained an embedded photograph of the device showing where to find the manufacturing date to determine if it is an affected device. The letter also informs the consignee they will be contacted by a GE HealthCare representative to arrange for the correction.
Initiated
2024-04-15
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, LA, MA, MD, MI, MO, NC, ND, NV, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Belgium, Brazil, Canada, Czech Republic, Finland, France, India, Israel, Italy, Japan, Malaysia, Malta, Poland, Singapore, Spain, Taiwan, Turkey, and United Kingdom.
Recall number
Z-2040-2024
Recall status
Open, Classified
CDRH record ID
207565
FDA classified
2024-06-07
Recall event number
94557

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗