Open original FDA record ↗Recall record
- FDA identifier
- Z-2041-2016
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc
- FDA record creation date
- Jun 21, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Exhibits a greater than 10% change in results in samples with biotin levels up to 10 ng/mL (41 nmol/L). Instructions for Use (IFU) states that specimens that have up to 10 ng/mL (41 nmol/L) of biotin
Source facts
- Action
- Siemens Healthcare sent an Urgent Medical Device Correction letter dated May 2, 2016, The letter identified the product, the problem and the action to be taken by the customer. Customers were instructed to continue use of the affected product, review the letter with their doctor, complete and return the Field Correction Effectiveness Check attached to the letter. If customers received any complaints of illness or adverse events associated wit the affected product they should immediately contact their local Siemens Customer Care Center or their local Siemens technical support representative. Customers with questions should contact their Siemens Customer Care Center or their local Siemens technical support representative. For questions regarding this recall call 508-668-5000.
- Initiated
- 2016-05-02
- Root cause
- Component design/selection
- Terminated
- 2018-02-22
- Distribution
- Worldwide Distribution - US (Nationwide) and Internationally to Albania Austria Belgium Bulgaria Burkina Faso Canada Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Polynesia Georgia Germany Great Britain Greece Guadeloupe Hungary Iceland Italy Kazakhstan Latvia Lithuania Luxembourg Macedonia Malta Martinique Netherlands Norway Poland Portugal Romania Russian Federation Serbia Slovakia Slovenia Spain Sweden Switzerland Tunisia Turkey United Arab Emirates and Uzbekistan.
- Recall number
- Z-2041-2016
- Recall status
- Terminated
- CDRH record ID
- 146020
- FDA record created
- 2016-06-21
- Recall event number
- 74036