Z-2041-2024: MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-150

Recall record Z-2041-2024 for Canon Medical System, USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2041-2024
Company in FDA record
Canon Medical System, USA, INC.
FDA classification date
Jun 9, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.

Source facts

Action
On 5/10/24, correction notices were sent to customers who were asked to do the following: 1) The replacement of the power cable securing parts that may cause smoke and fire, with parts made of materials with a higher degree of flame retardance, is now being released. 2) A firm representative will contact you to schedule a time to replace the power cable securing parts. 3) If any abnormalities such as unusual smells or smoke are observed before the work, immediately stop using the system and contact your service representative. 4) The parts required to address the defect are expected to be available no earlier than June 2024. 5) Complete and return the reply form via DocuSign or via email to pbiggins@us.medical.canon For questions, please contact our InTouch team at intouchdispatch@us.medical.canon
Initiated
2024-05-10
Root cause
Device Design
Distribution
US: PA, SC, IA, AL, MI, AR, SD, TX, ID, KY, NY, MN, MD, OH, WI, NV, CT, AZ, IL, OK, CA, FL, GA, OR, IN, NJ, CO, MT, KS, MO, WY, VA, VT, LA, NH, WV, TN, PR
Recall number
Z-2041-2024
Recall status
Open, Classified
CDRH record ID
207724
FDA classified
2024-06-09
Recall event number
94609

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗