Skip to content

Public FDA record

Z-2041-2025: HardyCHROM CRE, selective and differential culture medium for antimicrobial susceptibility testing, Produ

Recall record Z-2041-2025 for Hardy Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2041-2025
Company in FDA record
Hardy Diagnostics
FDA classification date
Jun 27, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Identified lots failed incoming quality control testing. Lots exhibited breakthrough growth of Klebsiella pneumoniae.

Source facts

Action
On May 6, 2025, the firm began notifying customers via "URGENT Medical Device Recall Notification" letters. Customers were instructed to discard any remaining stock and to destroy it according to local regulations. If product was further distributed, notify the consignees who received the product. Please retain this letter or post the letter where the product is stored to ensure the users are aware of this defect. For replacements or credit, please contact the firm's Technical Services Department at (800) 266-2222, option 2 or via email at TechnicalServices@HardyDiagnostics.com. Replacements will be sent out at no-charge to your laboratory or a credit will be applied to your invoice upon request. If you have any questions, call 805-346-2766 or email: malmberga@hardydiagnostics.com.
Initiated
2025-05-06
Root cause
Nonconforming Material/Component
Distribution
US domestic distribution to the following states: AZ, CA, CO, FL, GA, IL, MD, ME, MI, MN, MO, NC, NE, NV, NY, PA, PR, TX, UT, VA, WI and WV.
Recall number
Z-2041-2025
Recall status
Open, Classified
CDRH record ID
213768
FDA classified
2025-06-27
Recall event number
96843

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗