Public FDA record

Z-2042-2023: DigitalDiagnost C90, Model No. 712034 and 712035

Recall record Z-2042-2023 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2042-2023
Company in FDA record
Philips North America Llc
FDA classification date
Jun 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a software login in issue that may prevent the user from logging in.

Source facts

Action
Consignees in the United States were notified via Urgent Medical Device Correction Letter that was distributed on May 26, 2023. US consignees were initially notified via email. Follow-up notifications via FedEx will follow. Philips Market Organizations are responsible for distributing letters outside of the US in accordance with local requirements. A Philips Field Service Engineer will visit impacted customer sites to perform a software upgrade to prevent the issue from recurring. Philips has requested customers perform the following: a) To prevent the system from becoming inoperable, do not reboot or log out of your system until Philips has implemented the software correction; b) Circulate this notice to all users of the device so that they are aware of the issue; and c) Complete and return the response form to Philips promptly and no later than 30 days from receipt.
Initiated
2023-05-19
Root cause
Software change control
Distribution
Distribution in US: AZ, GA, IN, IA, MN, NE, NH, OH, TX OUS distribution to Australia, Chile, China, Malta, Namibia, New Zealand, Oman, Panama, Qatar, Suriname, United Kingdom
Recall number
Z-2042-2023
Recall status
Open, Classified
CDRH record ID
200971
FDA classified
2023-06-30
Recall event number
92462

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗