Open original FDA record ↗Recall record
- FDA identifier
- Z-2043-2016
- Company in FDA record
- Fujifilm Medical Systems U.S.A., Inc.
- FDA record creation date
- Jun 21, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Synapse cannot display image files, DICOM SR files, and/or Annotation files. The "Image Not Loaded" message displays instead
Source facts
- Action
- The firm, FujiFilm, sent an "MEDICAL DEVICE CORRECTION" letter dated May 10, 2016 via Federal Express to its customers. The letter describes the product, problem and actions to be taken.The customers were instructed to: Extend the Ad hoc deletion retention period or discontinue use of Ad Hoc Deletion, and complete and submit the attached Customer Response Form to Fujifilm via fax to (203)-251-7863. Affected systems will be corrected via on-site visits by FMSU Customer Service Engineers upgrading to Version 4.4.100. Questions or concerns, contact the FUJI FILM Customer Support Center at 1- 888-FUJIMED Option #1 .
- Initiated
- 2016-05-10
- Root cause
- Software design
- Terminated
- 2019-02-20
- Distribution
- US Distribution to states of: CA, FL, KS, NE, NY, and PA.
- Recall number
- Z-2043-2016
- Recall status
- Terminated
- CDRH record ID
- 147292
- FDA record created
- 2016-06-21
- Recall event number
- 74372