Z-2043-2016: Fujifilm Synapse PACS software version 4.4.000, Fujifilm Synapse PACS software version 4.4.001, Fujifilm

Recall record Z-2043-2016 for Fujifilm Medical Systems U.S.A., Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2043-2016
Company in FDA record
Fujifilm Medical Systems U.S.A., Inc.
FDA record creation date
Jun 21, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Synapse cannot display image files, DICOM SR files, and/or Annotation files. The "Image Not Loaded" message displays instead

Source facts

Action
The firm, FujiFilm, sent an "MEDICAL DEVICE CORRECTION" letter dated May 10, 2016 via Federal Express to its customers. The letter describes the product, problem and actions to be taken.The customers were instructed to: Extend the Ad hoc deletion retention period or discontinue use of Ad Hoc Deletion, and complete and submit the attached Customer Response Form to Fujifilm via fax to (203)-251-7863. Affected systems will be corrected via on-site visits by FMSU Customer Service Engineers upgrading to Version 4.4.100. Questions or concerns, contact the FUJI FILM Customer Support Center at 1- 888-FUJIMED Option #1 .
Initiated
2016-05-10
Root cause
Software design
Terminated
2019-02-20
Distribution
US Distribution to states of: CA, FL, KS, NE, NY, and PA.
Recall number
Z-2043-2016
Recall status
Terminated
CDRH record ID
147292
FDA record created
2016-06-21
Recall event number
74372

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗