Z-2043-2023: Access Hybritech p2PSA reagent, REF B03704 and REF A49752.

Recall record Z-2043-2023 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2043-2023
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Jun 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reagent lot numbers include an insufficient concentration of blocking reagent.

Source facts

Action
The recalling firm issued letters to U.S. consignees dated 5/15/2023 via first class mail and email on 5/17/2023. The letter described the issue, impact, and the action to be taken. The consignee was instructed to discontinue using the reagent lots that were listed and discard all remaining reagent packs from the lots. The letter was to be shared with the consignee laboratory and/or medical director regarding the need to review previous patient test results. The Beckman Coulter representative was to be contacted for replacement product requests. If the product was forwarded to another laboratory, the consignee is to provide them a copy of the letter. The consignee is requested to respond within 10 days of receipt of the letter either electronically or manually by completing the enclosed Response Form. The letter for the OUS consignees was dated 5/16/2023 and was similar to the letter for the U.S. consignees.
Initiated
2023-05-17
Root cause
Nonconforming Material/Component
Distribution
Distribution was made to CA, CO, CT, FL, GA, IL, IN, MA, MI, NC, NE, NJ, NY, OH, SC, TN, TX, UT, and WA. There was government distribution but no military distribution. Foreign distribution was made to Australia, Austria, Belgium, Brazil, China, Czech Republic, France, Germany, Hong Kong, India, Italy, Japan, Korea, Lithuania, Macao, Mexico, Portugal, Russia, Serbia, Singapore, Slovakia, Spain, Switzerland, Taiwan, and Vietnam.
Recall number
Z-2043-2023
Recall status
Open, Classified
CDRH record ID
200526
FDA classified
2023-06-30
Recall event number
92369

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗