Use of the Blood Pressure Graph located on the Crit-Line screen of the 2008T Hemodialysis Machine may result in a frozen screen, possibly causing the blood pump to stop.
Source facts
Action
Fresenius Medical Care Renal Therapies Group, LLC issued Recall Letter issued 5/24/16 via certified mail. Clinics actively using the CLiC option have been verbally communicated to stop using the CLiC option at . The other clinics that did not activate the CLiC option have been verbally communicated to not activate the CLiC option at this time. A Software update to all 2008T Hemodialysis machines will be required.
A reply form is included as part of the customer communication with instructions to fill out and return via fax or scanned/emailed copy.
For technical Inquiries, please contact Technical Services at 1-800-227-2572. Additional medical concerns or questions, please contact the Medical Information and Communication line at 855-616-2309 or website www.fresenius-medinfo.com.
Initiated
2016-05-24
Root cause
Software design
Terminated
2021-04-30
Distribution
Distributed US (nationwide)
Recall number
Z-2044-2016
Recall status
Terminated
CDRH record ID
146919
FDA record created
2016-06-21
Recall event number
74241