Reason stated in the source
An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
Source facts
- Action
- The firm notified their consignees by e-mail on 09/12/2014. Customers were informed of the reason for the recall, when the problem occurs, and what actions to take. Customers can perform a simple test by checking to see if there is a mismatch. Also, it is recommended that customers using BOLD and DTI, or other functionalities or workflows involving MPR switch to nordicBrainEx. Questions should be directed to NordicNeuroLab Customer Service at customerservice@nordicneurolab.com or 1-262-337-2909.
- Initiated
- 2014-09-12
- Root cause
- Software design
- Terminated
- 2020-10-06
- Distribution
- Worldwide Distribution -- USA, Australia, Denmark, China, France, Japan, Malaysia, Netherlands, USA, Poland, Sweden, Spain, Russia, India, Norway, Singapore, South Korea, United Kingdom, and Slovenia.
- Recall number
- Z-2044-2018
- Recall status
- Terminated