Z-2044-2018: nordicICE 2.3.14 Image processing software package used by trained professionals, including physicians

Recall record Z-2044-2018 for NordicNeuroLab AS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2044-2018
Company in FDA record
NordicNeuroLab AS
FDA record creation date
May 31, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.

Source facts

Action
The firm notified their consignees by e-mail on 09/12/2014. Customers were informed of the reason for the recall, when the problem occurs, and what actions to take. Customers can perform a simple test by checking to see if there is a mismatch. Also, it is recommended that customers using BOLD and DTI, or other functionalities or workflows involving MPR switch to nordicBrainEx. Questions should be directed to NordicNeuroLab Customer Service at customerservice@nordicneurolab.com or 1-262-337-2909.
Initiated
2014-09-12
Root cause
Software design
Terminated
2020-10-06
Distribution
Worldwide Distribution -- USA, Australia, Denmark, China, France, Japan, Malaysia, Netherlands, USA, Poland, Sweden, Spain, Russia, India, Norway, Singapore, South Korea, United Kingdom, and Slovenia.
Recall number
Z-2044-2018
Recall status
Terminated
CDRH record ID
159622
FDA record created
2018-05-31
Recall event number
78463

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗