Skip to content

Public FDA record

Z-2044-2020: SELECTSILICONE" 100% SILICONE FOLEY CATHETER 2-WAY 10 Fr 3 ml - Product Usage: Medline Silicone Foley cat

Recall record Z-2044-2020 for Medline Industries Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2044-2020
Company in FDA record
Medline Industries Inc
FDA record creation date
May 18, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Loose silicone particulate was found to be present on the shaft of the silicone catheters.

Source facts

Action
Medline Industries, Inc. sent formal letters with response forms to each consignee affected via 1st class mail on May 28, 2019. Each consignee was also provided with a response form to return via fax or email. The recalled product that is will be quarantined at the corporate office located at Three Lakes Drive, Northfield, IL 60093-2753 pending destruction.
Initiated
2019-05-28
Root cause
Manufacturing material removal
Terminated
2023-01-30
Distribution
US Nationwide distribution including in the states of CO, IL, KY, MO, NJ, OR, TN, TX, VA, WA.
Recall number
Z-2044-2020
Recall status
Terminated
CDRH record ID
179415
FDA record created
2020-05-18
Recall event number
84791

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗