Skip to content

Public FDA record

Z-2044-2021: Toshiba Vantage Tiitan 1.5T with Standard gradient, Premium Open-Bore MRI System, REF: MRT-1504 - Product

Recall record Z-2044-2021 for Canon Medical System, USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2044-2021
Company in FDA record
Canon Medical System, USA, INC.
FDA record creation date
Jul 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.

Source facts

Action
On May 25,2021, Canon issued a "URGENT: MEDICAL DEVICE CORRECTION" via USPS First Class Return Receipt. On June 24,2021, Canon sent an updated recall communication via USPS First Class Return Receipt to affected consignees to provide additional information. In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1. Your local Canon Service person will schedule time to replace the applicable part with a part that meets specifications. At the same time an inspection of the gantry will be performed. 2. Until such time that the modification work is provided by your service representative if any smoke or abnormal odor are observed, immediately stop using the system and contact your service representative. 3. Please share this information with all users and reviewing radiologists as well as the clinical engineering or biomedical group at your facility. 4. Please complete the attached form that is attached and return by either fax, to the toll-free number shown at the top of the form, or by via email to RAffairs@us.medical.canon. This form is a record required by FDA for verification that this notification has been received. 5. If you have any questions, please feel free to contact Terry Schultz Regulatory Affairs Manager at (800) 421-1968 or your local Canon Medical Systems service representative at (800) 521-1968.
Initiated
2021-05-21
Root cause
Under Investigation by firm
Distribution
US Nationwide distributions in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Recall number
Z-2044-2021
Recall status
Open, Classified
CDRH record ID
187690
FDA record created
2021-07-09
Recall event number
88014

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗