Skip to content

Public FDA record

Z-2045-2016: Myriad handpiece Tissue morcellator

Recall record Z-2045-2016 for Nico Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2045-2016
Company in FDA record
Nico Corp.
FDA record creation date
Jun 21, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This recall has been initiated due to a defective component within the handpiece which will cause the handpiece to fail. The failure will result in the inability to cut tissue with the handpiece, rendering it non-functional. Should this failure occur, a replacement Myriad handpiece will be required. If a replacement handpiece is not available, alternate instrumentation for tissue removal will be necessary.

Source facts

Action
NICO sent an Urgent: NICO Myriad Handpiece Recall Notification on February 24, 2014, to all affected customers requesting return of the affected product. The recall notification included a description of the reason for the recall, affected product, customer responsibilities, and instructions for responding to the formal recall notification. If you have any questions regarding this matter, please call (317) 660-7118, ext. 104.
Initiated
2014-02-24
Root cause
Device Design
Terminated
2016-06-28
Distribution
US Distribution to the states of : OR, IL, IA, NC, TX, CA, GA, OH, WI, AR, MI, MD, IN.
Recall number
Z-2045-2016
Recall status
Terminated
CDRH record ID
146888
FDA record created
2016-06-21
Recall event number
74191

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗