Open original FDA record ↗Recall record
- FDA identifier
- Z-2045-2018
- Company in FDA record
- NordicNeuroLab AS
- FDA record creation date
- Jun 1, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is an error related to the relative geometry between fiber tracts in a fiber group and images. The error occurs in the following situations: A group of fibers has been selected using the VOI functionality and put into a fiber group. This is done with images series A visualized in the 3D viewer. Then a new images series B is visualized in the 3D viewer. If image series A and B have the same geometrical resolution (pixel sizes and slice distances), the coregistration matrix taking A to B will not be applied to the fiber group, and therefore the fiber group will in general not be positioned correctly on B in the 3D viewer. Furthermore, if the fiber group is exported as a new image series the fibers could be misplaced, and similarly if the fibers are visualized in the MPR on B, the positioning will not be correct. The misalignment will be equal to the rotations/translations necessary to align A with B.
Source facts
- Action
- The firm distributed the recall notification letter by email. The letter identified the affected product, problem and the actions to be taken. For questions contact NordicNeurolab Customer Service or if you would like to arrange a return customerservice@nordicneurolab.com.
- Initiated
- 2012-12-21
- Root cause
- Software design
- Terminated
- 2020-09-16
- Distribution
- Worldwide Distribution - US Nationwide - Saudi Arabia, Brazil, Sweden, Spain, Russia, Belgium, Norway, South Korea, Japan
- Recall number
- Z-2045-2018
- Recall status
- Terminated
- CDRH record ID
- 164721
- FDA record created
- 2018-06-01
- Recall event number
- 80162