Skip to content

Public FDA record

Z-2045-2023: wDR 2.2 Mobile Digital Diagnostic X-Ray System

Recall record Z-2045-2023 for SEDECAL SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2045-2023
Company in FDA record
SEDECAL SA
FDA classification date
Jun 30, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a software login in issue that may prevent the user from logging in.

Source facts

Action
A recall notification letter was mailed to customers in the United States by the US distributor, Philips, beginning May 23, 2023. Customers are advised to take the following actions: 1) To prevent the system from becoming inoperable, do not reboot or log out of your system until Philips has implemented the software correction. 2) Circulate this notice to all users of the device so that they are aware of the issue. 3) Complete and return the response form to Philips promptly and no later than 30 days from receipt. A Philips representative will contact consignees to schedule a time for a Field Service Engineer to visit the site and implement a software solution to resolve the issue.
Initiated
2023-05-23
Root cause
Software change control
Distribution
Distribution throughout United States including Puerto Rico and US Virgin Islands OUS distribution to Albania, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Czech Republic, Dominican Republic, Egypt, Finland, France, Germany, Hungary, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia,Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Martinique, Mexico, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Suriname, Sweden, Switzerland, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam, and Zambia
Recall number
Z-2045-2023
Recall status
Open, Classified
CDRH record ID
200975
FDA classified
2023-06-30
Recall event number
92464

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗