Skip to content

Public FDA record

Z-2046-2026: Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number

Recall record Z-2046-2026 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2046-2026
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
May 1, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.

Source facts

Action
On 04/02/2026, the firm sent via 2-day mailing service an "Urgent Medical Device Correction" letter to customers informing them that Intuitive identified that the screws on the cannula mount of the USM from a specific population are susceptible to breaking over time. When the cannula mount experiences a breakage of all six screws, the mount can separate from the rest of the USM. Breakage of any fewer than six screws will not result in cannula mount separation, or impact to clinical performance. How to recognize issue/failure: Screw breakage or cannula mount loosening may be observable as visible or tactile separation of the cannula mount from the body of the USM. Prior to a procedure, closely inspect the joint between the cannula mount and insertion axis that is joined by the screws in Figure 1 in the Urgent Medical Device Correction Letter. Manipulate the cannula mount joint by hand to ensure there is no movement or looseness. If looseness or separation is observed, discontinue use of the USM and contact representative to schedule a site visit. If all six screws were to break intra-operatively, the user may also observe imprecise or unintended instrument tip motion. If observed imprecise or unintended instrument tip motion, inspect the cannula mount for looseness or separation and if present, discontinue use of the USM and contact representative to schedule a site visit. Customers are instructed to: 1. This notice needs to be passed on to all those who need to be aware within your organization or functions where the potentially affected devices have been transferred. 2. Retain a copy of letter, place with your affected system, and keep a copy of the acknowledgement form for your files. For Questions: North/South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com. Europe/Middle East/Asia/Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or eucs@intusurg.com South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or support
Initiated
2026-04-02
Root cause
Nonconforming Material/Component
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
Recall number
Z-2046-2026
Recall status
Open, Classified
CDRH record ID
219087
FDA classified
2026-05-01
Recall event number
98564

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗