Z-2047-2025: Bravo CF capsule delivery device, designed to place and attach a pH monitoring capsule to the esophagus u

Recall record Z-2047-2025 for Given Imaging Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2047-2025
Company in FDA record
Given Imaging Ltd.
FDA classification date
Jul 10, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

It has been determined that misapplied adhesive on the Bravo CF capsule delivery device may lead to its malfunction. Specifically, the misapplied adhesive may prevent the capsule from attaching to the patient's esophagus or detaching from the delivery device.

Source facts

Action
An URGENT: MEDICAL DEVICE RECALL notification letter was sent to customers beginning on 6/3/25. Actions: "Immediately identify and quarantine all unused FGS-0635 (Bravo CF capsule delivery device, 5-pk)and FGS-0636 (Bravo CF capsule delivery device, 1-pk). See Attachment A and Attachment B for affected lot numbers. "Return all unused, affected product(s) to Medtronic for replacement/credit as described on the Customer Confirmation Form. "Please complete and e-mail the enclosed Customer Confirmation Form to rs.gmbmitgfca@medtronic.com, even if you do not have unused inventory. "Pass on this notice to all those who need to be aware within your organization or to any organization where the affected product has been transferred or distributed. Local contact details: Adverse reactions or quality problems experienced with this product should be reported to the FDA and Medtronic: "Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail) or call FDA (800) FDA-1088 "Call Medtronic at 800-448-3644, Option 3 An Update to URGENT: MEDICAL DEVICE RECALL notification letter dated 6/16/25 was sent to customers. Medtronic recently notified you of a voluntary recall of specific lots of the Bravo CF capsule delivery device. In the original notification letter sent on 03 June 2025 we separated impacted lot numbers by product number. Please note that the Lot numbers listed in Appendix A below can be found on both the 5-pk FGS-0635 outer label and on the 1-pk FGS-036 individual units. Customers should check the full list of Lot numbers listed in the appendices when inspecting inventory. A revised Appendix A is attached to this letter. If you have already completed product inspection and returned the Customer Confirmation form, please re-inspect your inventory against the full list of impacted Lot numbers and re-submit the form to: rs.gmbmitgfca@medtronic.com
Initiated
2025-06-03
Root cause
Process control
Distribution
US Nationwide. Global Distribution.
Recall number
Z-2047-2025
Recall status
Open, Classified
CDRH record ID
214251
FDA classified
2025-07-10
Recall event number
97029

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗