Z-2048-2021: Hang&Go (kit for hyperthermic perfusion), Reference Code R9900120

Recall record Z-2048-2021 for RanD S.r.l.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2048-2021
Company in FDA record
RanD S.r.l.
FDA record creation date
Jul 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The sterilization contractor informed the company of a quality issue possibly affecting the sterilization process of a certain number of products. It was determined that some lots may not be sterile.

Source facts

Action
An Urgent Medical Device Recall notification letter dated 4/30/21 was sent to customers. Reason for the Voluntary Recall: Our sterilization contractor informed us about a quality issue possibly affecting the sterilization process of a certain number of RAND products, thus leading us to open an immediate investigation. We have determined that the lots mentioned in the table below may be non-sterile. Risk to Health: The exposure to non-sterile products may result in an increased risk of patient infection. However we have not received any adverse event report regarding patient infection events associated with this field safety notice, so far. No specific follow up activities are required if the products have already been used. Actions to be taken by the Customer/User: Please proceed as follows: 1. Please check if you still have in your stock products belonging to the lot numbers below. Please refer to the reference code (REF) and the lot number (LOT) on the product label. 2. If you identify affected product, cease use immediately and put these products in quarantine. Please be sure that they are properly and clearly identified as on-hold to prevent their unintentional use. 3. Please fill in the form on the last page and return it to cristina.vaccari@rand-biotech.com as soon as possible, but no later than May 30, 2021. The form must be completed and returned even if you do not have affected product on hand. 4. If you identified non-conforming products in your inventory, you will receive further instructions on product disposition (return them to RAND or destroy them). Type of Action by the Company: No alternative products can be used, because only RanD sterile medical devices can be used in combination with the Performer HT equipment. RanD S.p.A. has already validated a different sterilization provider. New sterile products will be available in 1 - 2 month time. Contact details: If you have any questions or concerns regarding this notification, d
Initiated
2021-04-30
Root cause
Material/Component Contamination
Terminated
2024-02-05
Distribution
The products were distributed to the following US states: FL, IL, TN, WI
Recall number
Z-2048-2021
Recall status
Terminated
CDRH record ID
187451
FDA record created
2021-07-09
Recall event number
87926

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗