Due to a design change in 2015, corrosion on the sample probe electrical contacts can occur and lead to a disturbance of the sample liquid level detection (LLD). When disturbance of the sample liquid level detection (LLD) occurs, the affected
sample probe may dip into the sample material deeper than intended. This may lead to carryover and medical risk cannot be excluded.
Source facts
Action
On March 21, 2017, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected consignees via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification.
Initiated
2017-03-21
Root cause
Component change control
Terminated
2017-10-02
Distribution
US Nationwide.Distribution
Recall number
Z-2049-2017
Recall status
Terminated
CDRH record ID
154407
FDA record created
2017-05-09
Recall event number
76854