Skip to content

Public FDA record

Z-2050-2020: Professional Urine Drug Control, Item code 88010 - Product Usage: The Alere Professional Cup Urine Drug C

Recall record Z-2050-2020 for Cardinal Health Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2050-2020
Company in FDA record
Cardinal Health Inc.
FDA record creation date
May 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Three products requiring refrigerated storage conditions were incorrectly stored outside of approved temperature ranges. Using these temperature abused products may result in misdiagnosis of the patient.

Source facts

Action
On April 20, 2020, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed that the product had been incorrectly stored and that there was a potential for misdiagnosis. Action Required: 1. CHECK all storage and usage locations to confirm whether you have any units of the affected product codes and lot numbers containing the labeling outlined in the letter. 2. SEGREGATE and QUARANTINE all on-hand affected product. 3. PLEASE RETURN the letter's acknowledgment form via facsimile (847-689-9101 or 614-652-9648) or email (GMB-FieldCorrectiveAction@cardinalhealth.com) and indicate the product code, lot and quantity of product youve quarantined or discarded. Please respond regardless of whether or not you have affected product. 4. NOTIFY any customers to whom you may have distributed, or forwarded product affected by this recall. Your notification to your customers may be enhanced by including a copy of this recall notification letter. 5. CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5:00pm EST: " Hospital800-964-5227 " Federal Government800-444-1166 " Distributor800-635-6021 " All other Customers888-444-5440
Initiated
2020-04-20
Root cause
Storage
Terminated
2020-07-23
Distribution
US Nationwide distribution NC, KS, and AL.
Recall number
Z-2050-2020
Recall status
Terminated
CDRH record ID
181118
FDA record created
2020-05-20
Recall event number
85509

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗