Skip to content

Public FDA record

Z-2050-2021: EcoFit cup cementless: Model(Catalog Number)/Description: 02200046 /EcoFit cup cementless 46mm dia incl

Recall record Z-2050-2021 for IMPLANTCAST GMBH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2050-2021
Company in FDA record
IMPLANTCAST GMBH
FDA classification date
Jul 9, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A risk exists during impacting of the acetabular cup into the bone the plugs may come loose from the cup, which could lead to operation duration extension or revision operation. A warning note is being added to instructions for use and surgical technique alerting users that the quality of bone in the patient (such as osteosclerosis) is an attributable factor.

Source facts

Action
On May 10, 2021 Urgent Safety Information notices were emailed to customers who were asked to take the following action: 1. Read this safety information carefully and make sure all relevant departments and officeholders are informed about its content. 2. Keep this safety information. 3. Fill in and return the reply form. Customers with additional questions were encouraged to contact Wendy Ryan-Kettler at WRyan@implantcast-na.com On 17 May, 2017, a follow-up Medical Device Correction notice was emailed to customers who were asked to take the following additional actions: 4. Destroy the affected (previous) surgical technique and instruction for use (IFU) and replace them with the updated versions that accompanied this safety information.
Initiated
2021-05-10
Root cause
Device Design
Terminated
2022-11-18
Distribution
U.S. Nationwide distribution in the states of FL and TX.
Recall number
Z-2050-2021
Recall status
Terminated
CDRH record ID
187774
FDA classified
2021-07-09
Recall event number
88047

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗