Open original FDA record ↗Recall record
- FDA identifier
- Z-2050-2021
- Company in FDA record
- IMPLANTCAST GMBH
- FDA classification date
- Jul 9, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A risk exists during impacting of the acetabular cup into the bone the plugs may come loose from the cup, which could lead to operation duration extension or revision operation. A warning note is being added to instructions for use and surgical technique alerting users that the quality of bone in the patient (such as osteosclerosis) is an attributable factor.
Source facts
- Action
- On May 10, 2021 Urgent Safety Information notices were emailed to customers who were asked to take the following action: 1. Read this safety information carefully and make sure all relevant departments and officeholders are informed about its content. 2. Keep this safety information. 3. Fill in and return the reply form. Customers with additional questions were encouraged to contact Wendy Ryan-Kettler at WRyan@implantcast-na.com On 17 May, 2017, a follow-up Medical Device Correction notice was emailed to customers who were asked to take the following additional actions: 4. Destroy the affected (previous) surgical technique and instruction for use (IFU) and replace them with the updated versions that accompanied this safety information.
- Initiated
- 2021-05-10
- Root cause
- Device Design
- Terminated
- 2022-11-18
- Distribution
- U.S. Nationwide distribution in the states of FL and TX.
- Recall number
- Z-2050-2021
- Recall status
- Terminated
- CDRH record ID
- 187774
- FDA classified
- 2021-07-09
- Recall event number
- 88047