Reason stated in the source
The Firm discovered a software error on the Accuson S2000 ultrasound systems affecting the measurement results in the patient report.
Source facts
- Action
- Siemens issued a Customer Safety Advisory Notification letter to all customers on March 15, 2011, regarding a software error on the ACUSON S2000 ultrasound systems affected measurement . They are in the process of releasing a software update that will correct this issue. Siemens recommends reviewing the patient report at the end of each study to confirm that the report contains only measurements relevant to the exam just performed. No patient injury has been reported. Further questions regarding this recall please call (650) 694-5398.
- Initiated
- 2011-03-15
- Root cause
- Software design
- Terminated
- 2012-06-18
- Distribution
- Worldwide Distribution including USA, states of AZ, VA, CA, NE, IL, MN, TN, FL, NM, and DC
- Recall number
- Z-2051-2011
- Recall status
- Terminated
- CDRH record ID
- 98764
- FDA classified
- 2011-04-22
- Recall event number
- 58231