Skip to content

Public FDA record

Z-2052-2026: Brand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module Model/Catalog Number: 9

Recall record Z-2052-2026 for Spacelabs Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2052-2026
Company in FDA record
Spacelabs Healthcare, Inc.
FDA classification date
May 1, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).

Source facts

Action
On 03/26/2026, the firm sent via email an URGENT MEDICAL DEVICE CORRECTION Letter to customers informing them that Spacelabs Healthcare became aware that for certain Multi-Parameter Command Modules (Model 91496) manufactured between 07/17/2025 to 08/05/2026 using a malfunctioning test equipment resulting in inadequate electrical leakage testing. Customers are instructed to: 1.Notify Please notify all relevant employees within your organization, including facilities you have distributed the products to, of this medical device correction notice and to cease use of the modules identified in Appendix B. 2.Acknowledge Please acknowledge receipt of this notice by either utilizing the URL/QR code listed in Appendix A or completing the paper acknowledgement form and returning it via one of the methods outlined in the form. 3.Segregate Identify and Quarantine the affected devices listed in Appendix B. 4.Report Contact your local Spacelabs Healthcare service representative to coordinate your device correction as soon as possible. To remedy the issue identified in this letter, Spacelabs Healthcare will be replacing the impacted devices with brand new units of the same specification. For questions - call Phone: 1-800-522-7025 Website: https://spacelabshealthcare.com/support/
Initiated
2026-03-26
Root cause
Process control
Distribution
Worldwide - US Nationwide di GA, KS, MI, OH, SC, TX, WA and the country of ARGENTINA.
Recall number
Z-2052-2026
Recall status
Open, Classified
CDRH record ID
219564
FDA classified
2026-05-01
Recall event number
98722

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗