Open original FDA record ↗Recall record
- FDA identifier
- Z-2054-2011
- Company in FDA record
- Smith & Nephew Inc
- FDA classification date
- Apr 22, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
During the manufacturing process for several batches of R3 Ceramic Liners, the titanium rings were pressed onto the ceramic component with a higher force than allowed by manufacturing specifications. This has the potential to result in lower than expected strength for the liners.
Source facts
- Action
- All affected Smith & Nephew Distributors were notified via email and the hospitals were notified by letter delivered by Fed-Ex on 3/11/2011. They were instructed to immediately cease distribution, quarantine affected product for return, if product further distributed, identify customers and immediately notify them of recall. Please contact Mr. David Archer, Group Director, at 701.396.2121, ext. 5373, if you have questions.
- Initiated
- 2011-03-11
- Root cause
- Employee error
- Terminated
- 2012-02-08
- Distribution
- AK, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OR, PA, SC, TN, TX, WA, and WI
- Recall number
- Z-2054-2011
- Recall status
- Terminated
- CDRH record ID
- 98987
- FDA classified
- 2011-04-22
- Recall event number
- 58324