Open original FDA record ↗Recall record
- FDA identifier
- Z-2054-2013
- Company in FDA record
- Stryker Craniomaxillofacial Division
- FDA classification date
- Aug 26, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Stryker received reports of the desiccant bag in the kit leaking silica gel granulate. This could increase exposure to anesthesia during surgical delay up to 60 minutes or inadvertent implantation of silica gel granulate which could cause foreign body reaction including inflammation and might lead to the need for revision surgery.
Source facts
- Action
- Stryker sent an Urgent Medical Device Recall Notices with attached Product Correction Bulletin dated, July 30, 2013, describing the product, problem and possible adverse events associated with the recall were sent to all affected customers. Customers were instructed to pass the notice and bulletin to all individuals who may be using the device, provided instructions on procedures which would help mitigate risk and return the customer response form to our Customer Service Team by email, fax or letter. Email: CMF-custserv@stryker.com Fax: 1-877-648-7114 Address: 750 Trade Centre Way, Suite 200 Portage, MI 49002 Customers responses will be tracked and non-responders will be contacted. Further questions please call (877) 534-2464.
- Initiated
- 2013-07-30
- Root cause
- Package design/selection
- Terminated
- 2014-04-17
- Distribution
- Worldwide Distribution - USA (nationwide) including the countries through Stryker Distribution: Austria, Belgium, Denmark, Estonia, Finland, France, Germany, Ireland, Italy, Latvia, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Serbia, Spain, Sweden, Switzerland, United Kingdom, Argentina, Australia, Brazil, Jordan, Kingdom of Saudi Arabia, Kuwait, Latin America (others), New Zealand, Russia, Singapore, South Africa, and United Arabian Emirates.
- Recall number
- Z-2054-2013
- Recall status
- Terminated
- CDRH record ID
- 120159
- FDA classified
- 2013-08-26
- Recall event number
- 65787