Open original FDA record ↗Recall record
- FDA identifier
- Z-2055-2014
- Company in FDA record
- Stryker Communications
- FDA classification date
- Jul 15, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During the use of the SHAPE Arm there is the potential for the monitor mount to come off of the SHAPE Arm if the lock washer and screw are not present.
Source facts
- Action
- Direct accounts in the US were notified by letter dated March 21, 2014, and asked to report any malfunctions to Stryker Communications immediately. In addition, Stryker Representatives will visit all US accounts to inspect all the product units on hand. If needed, Representatives will check that black, screw/lock washers supplied by Stryker are secured in place and will replace any that are not the typed furnished by Stryker. All units will be updated with a yellow, warning sticker on the monitor bracket which reads: "Failure to properly install the screw and lock washer below can result in the attached component/equipment falling." Stryker Communications' foreign distributors will oversee notification, inspection, and corrections of their accounts and report corrections and other findings to Stryker Communications.
- Initiated
- 2014-03-17
- Root cause
- Process control
- Terminated
- 2015-05-15
- Distribution
- Worldwide Distribution. US nationwide including Puerto Rico, China, Philippines, Mexico, Korea, Japan, Colombia, Chile, Canada, Brazil, and Australia
- Recall number
- Z-2055-2014
- Recall status
- Terminated
- CDRH record ID
- 128239
- FDA classified
- 2014-07-15
- Recall event number
- 67813