The device is a Left 8 hole plate but is incorrectly labeled and etched as a Right device.
Source facts
Action
On 15-June-2012, a follow-up, URGENT Recall of MIS-BRANDED PediLoc Locking Plate e-mail was sent to all Distributors and direct representatives to provide them with a written notification of the voluntary recall. The e-mail identified the affected product and the reason for the recall. It also instructed them to quarantine all identified devices from the affected lot number immediately. A certified Voluntary Recall notice was also sent as a follow up along with a return prepaid pouch for the return of the recalled device(s) and Mandatory Reply Form.
Initiated
2012-06-14
Root cause
Error in labeling
Terminated
2013-04-30
Distribution
Worldwide Distribution -- USA, including states of IN, GA, MI, FL, MO, OH, and SC and country of Australia.
Recall number
Z-2057-2012
Recall status
Terminated
CDRH record ID
110311
FDA classified
2012-07-20
Recall event number
62351