Open original FDA record ↗Recall record
- FDA identifier
- Z-2057-2015
- Company in FDA record
- Hamilton Medical, Inc.
- FDA classification date
- Jul 10, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device.
Source facts
- Action
- *** Please note *** This is not a new recall. An error occurred, and this recall was originally classified as a class I on July 10, 2015. No further action is required. Hamilton Medical AG notified their customers with a letter dated February 19, 2015 and had a fix (SW 2.42) that was available as of February 20th 2015. Hamilton Medical proceeded to send out notification letters the 24th of March 2015 once they compiled the complete and validated list of customers that had the affected software. Customers with questions were instructed to call 800-426-6331, ext.215.
- Initiated
- 2015-03-24
- Root cause
- Device Design
- Terminated
- 2015-12-23
- Distribution
- Nationwide Distribution
- Recall number
- Z-2057-2015
- Recall status
- Terminated
- CDRH record ID
- 136935
- FDA classified
- 2015-07-10
- Recall event number
- 71234