Open original FDA record ↗Recall record
- FDA identifier
- Z-2057-2019
- Company in FDA record
- CooperSurgical, Inc.
- FDA record creation date
- Jul 23, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The products' cut, coagulate, or blend function may not operate with the use of the Foot Pedal.
Source facts
- Action
- The firm notified customers of the recall via Urgent Medical Device Recall letters dated June 18, 2019. Customers were informed that the device may fail to cut, coagulate, or blend tissue with use of the Foot Pedal. Serious injuries could occur due to this failure mode. Customers were instructed to do the following: - Use the Hand-Switch Pencil and Cord (Part Numbers: 395310/LP50101) to activate the cut, coagulation, or blend functions by hand control instead of foot action using the Foot Pedal. - Complete the acknowledgement and receipt form and return to the firm. If you have additional questions, please contact a CooperSurgical Product Surveillance representative at 203.601.5200 Ext. 3300 or email recall@coopersurgical.com.
- Initiated
- 2019-06-18
- Root cause
- Device Design
- Terminated
- 2020-06-15
- Distribution
- Nationwide domestic distribution. International distribution to Canada and Costa Rica.
- Recall number
- Z-2057-2019
- Recall status
- Terminated
- CDRH record ID
- 173570
- FDA record created
- 2019-07-23
- Recall event number
- 83100