Open original FDA record ↗Recall record
- FDA identifier
- Z-2058-2013
- Company in FDA record
- Synthes USA HQ, Inc.
- FDA classification date
- Aug 26, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Recall was initiated due to the possibility that the outer pouch was compromised. The product is packaged in two pouches; a sterile, internal pouch within an external pouch. It is the outer, external pouch that may be compromised.
Source facts
- Action
- The firm initiated their recall of this product on November 2, 2012 by sending an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter to all consignees. The letter explained the problem and gave instructions for consignees to examine their inventory and urged consignees to return the identified devices. The letter stated If returning the product would adversely impact their ability to provide necessary medical care to patients, they should re-sterilize the product per Pre-Vacuum Sterilization instructions contained in the Instruction for Use.
- Initiated
- 2012-11-02
- Root cause
- Packaging change control
- Terminated
- 2016-09-27
- Distribution
- Nationwide distribution.
- Recall number
- Z-2058-2013
- Recall status
- Terminated
- CDRH record ID
- 119782
- FDA classified
- 2013-08-26
- Recall event number
- 65631