Z-2058-2017: Medi-Therm Hyper/Hypothermia System, MTA7900

Recall record Z-2058-2017 for Stryker Medical Division of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2058-2017
Company in FDA record
Stryker Medical Division of Stryker Corporation
FDA record creation date
May 10, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The water may reheat too quickly if the unit is in "Automatic" mode, switched to "Manual" mode to view water temperature, and then switched back to "Automatic" mode.

Source facts

Action
Stryker sent an URGENT: MEDICAL DEVICE CORRECTION notification dated March 29, 2017, were sent to the affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Actions needed 1. Locate the units listed on the attached business reply form and identify the address where they can be serviced. 2. In the interim, until the service can be performed, there are additional actions that users can and should take to ensure that the product operates safely. a. Attach the enclosed warning tags to the units listed in this notice. b. For units that do not have the water temperature displayed, advise staff not to switch to manual mode in order to display the water temperature. 3. Return the enclosed business reply form to confirm receipt of this notification by fax 269 488 8691 or email productfieldaction@stryker.com. 4. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the modification/repair of your Medi-Therm unit. 5. If you have loaned or sold any of the products listed in this letter, please forward a copy of this notice to the new users and advise us of their new location in the space provided on the business reply form. 6. If you have disposed of any of these units and they are no longer in use, please, advise us of their obsolescence by providing us with their serial number in the space provided on the business reply form. For further questions, please call (800) 327-0770.
Initiated
2017-03-29
Root cause
Software design
Terminated
2017-09-18
Distribution
USA (nationwide) Distribution and Internationally to Colombia.
Recall number
Z-2058-2017
Recall status
Terminated
CDRH record ID
154824
FDA record created
2017-05-10
Recall event number
77014

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗