Z-2058-2018: Cell Observer SD, DirectFRAP, Laser TIRF 1, Laser TIRF 2, Laser TIRF3, LightSheet, Elyra, and multiple L

Recall record Z-2058-2018 for Zeiss, Carl Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2058-2018
Company in FDA record
Zeiss, Carl Inc
FDA classification date
Jun 9, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In certain eyepiece configurations the laser shutter may not close completely, permitting reflected beams that may be greater than the Class I limit.

Source facts

Action
1.Your firm will send an initial email to all known users advising them of the issue, the risk involved and associated mitigations until the units can be evaluated by Carl Zeiss service representatives, 2.Your firm will send a second written communication to each first line purchaser and to each known user site with the notice of action and warning to discontinue use of the component until it has been verified as defective or properly functioning, and 3.Your firm will dispatch factory trained service representatives to perform verification tests and to replace defective units as necessary, free of charge. CDRH approves the CAP subject to the following conditions: For further questions, please call (914) 747-1800.. 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Corrective actions will be provided at no cost to the purchasers and completed by March 1, 2019.
Initiated
2018-05-08
Root cause
Radiation Control for Health and Safety Act
Terminated
2020-04-21
Distribution
USA (nationwide)
Recall number
Z-2058-2018
Recall status
Terminated
CDRH record ID
164575
FDA classified
2018-06-09
Recall event number
80115

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗