Z-2059-2008: ABBOTT Clinical Chemistry Phenytoin, List Number: 1E07-20, is used for the quantitation of phenytoin in

Recall record Z-2059-2008 for Abbott Laboratories Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2059-2008
Company in FDA record
Abbott Laboratories Inc.
FDA classification date
Sep 16, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

An Abbott investigation has determined that the phenytoin reagent does not maintain an onboard stability of 28 days as stated in the phenytoin package insert. In addition, the 14 day calibration interval is also not being met. Further investigation is required to determine the root cause of this issue. Until the cause has been determined and a corrective action defined, specific instructions ar

Source facts

Action
On February 15, 2008 A Product Correction letter with instrument specific attachments was provided to all Clinical Chemistry Phenytoin customers that have received the affected lots. A customer reply form was included in all US customers letter. The letter instructed customers to follow the attached instrument specific actions as described in the attachments (Attachment A: ARCHITECT cSYSTEM, Attachment B: AEROSET). Customers were also instructed to forward a copy of the Product Correction letter to any other laboratories to which they may have forwarded the affected product. If you have questions, U.S. customers should call Customer Support at 1-800-4ABBOTT and customers outside of the U.S. should contact your local customer support.
Initiated
2008-02-15
Root cause
Other
Terminated
2012-03-23
Distribution
Nationwide and Worldwide to: Australia, Canada, Chile, Germany, Hong Kong, New Zealand, Panama, Singapore, Uruguay, Singapore
Recall number
Z-2059-2008
Recall status
Terminated
CDRH record ID
70930
FDA classified
2008-09-16
Recall event number
48173

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗