Open original FDA record ↗Recall record
- FDA identifier
- Z-2059-2019
- Company in FDA record
- Zyno Medical LLC
- FDA record creation date
- Jul 23, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified
Source facts
- Action
- On May 29, 2019 Zyno issued letter to the Distributor advising of the problem and health risk and action to take: remove all recalled products with codes, cease use and immediately and return them to Zyno Medical for replacement from unaffected lots. Call or email for instructions on the return of products: (508) 907-7772 (Monday to Friday, 9:30 a.m. to 5 p.m. EST) or wagner.Cruz@zvnomed.com to acknowledge receiving this recall notice. Zyno issued a follow-up recall letter on 6/10/19 via email to the Distributor to issue to the user/customer level.
- Initiated
- 2019-05-29
- Root cause
- Process control
- Terminated
- 2020-06-22
- Distribution
- MA
- Recall number
- Z-2059-2019
- Recall status
- Terminated
- CDRH record ID
- 173367
- FDA record created
- 2019-07-23
- Recall event number
- 83026