Open original FDA record ↗Recall record
- FDA identifier
- Z-2059-2023
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA classification date
- Jun 30, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Fenestrated Graspers have a manufacturing issue has the potential to cause breakage on both upper and/or lower grips, which could lead to stainless steel fragments falling into the patient that may be either detected or undetected.
Source facts
- Action
- On 6/12/23 Field Safety Notices were mailed and emailed to customers asking them to do the following: 1) Locate and return affected devices. 2) Notify all surgeons and users of the device about the contents of the notice. 3) Complete and return the acknowledgement form to Recalls@intusurg.com If you need further information or support concerning this Medical Device Notification, please contact your Clinical Sales Representative or contact Customer Service at: - North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customerservice@intusurg.com. - Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or ics@intusurg.com - South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) - Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST) - India: +1-800-103-6952 (9 AM to 6 PM IT) - Taiwan: +0800-86-8181 (9 AM to 6 PM CT)
- Initiated
- 2023-06-12
- Root cause
- Process control
- Distribution
- Worldwide - US Nationwide distribution in the states of PR, FL, MN, PA, AZ, OH, IL, NC, GA, KS, CA, HI, MD, OR, SC, AK, NY, NE, IA, WA, OK, ND, UT, MS, AR, WI, MI, TX, TN, IN, NJ, VA, SD, CO, KY, AL, MO, LA, DE, MA, MT, WV, ID, CT, NV, NH, ME, WY, DC, RI, NM, VT and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Luxembourg, Mexico, Monaco, Netherlands, Norway, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.
- Recall number
- Z-2059-2023
- Recall status
- Open, Classified
- CDRH record ID
- 200883
- FDA classified
- 2023-06-30
- Recall event number
- 92418