Open original FDA record ↗Recall record
- FDA identifier
- Z-2060-2015
- Company in FDA record
- Zimmer, Inc.
- FDA classification date
- Jul 13, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A single distributed ZNN Greater Trochanter Femoral Nail may have been dented. Potential for fatigue of implant prior to sufficient fracture healing, which may result in need for implant removal and revision of fracture fixation to assure proper healing and avoid fracture malunion or nonunion.
Source facts
- Action
- On 6/17/2015, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Managers with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-877-946-2761. Hours of operation are Monday through Friday, 8 a.m. through 5 p.m. EST.
- Initiated
- 2015-06-17
- Root cause
- Process control
- Terminated
- 2016-03-15
- Distribution
- Worldwide distribution. US states of NC and AK; Taiwan, Japan, and Germany.
- Recall number
- Z-2060-2015
- Recall status
- Terminated
- CDRH record ID
- 138114
- FDA classified
- 2015-07-13
- Recall event number
- 71530