Reason stated in the source
Intravenous fluid may leak at the CLAVE secondary port of Hospira PlumSet IV administration sets.
Source facts
- Action
- Hospira sent an URGENT MEDICAL DEVICE CORECTION letters dated September 9, 2013 to customers. The letter identified the affected product, problem and actions to be taken. the letter included additional instructions on the proper handling/use of the CLAVE devices. The letters request that all incidents of separation or breakage of CLAVE from the secondary port of the cassette be reported to Hospira by calling 1-800-441-410 (Monday - Friday, 8:00 AM - 5:00 PM CST) or by e-mail at ProductComplaintPP@hospira.com. Customers are further requested to complete and return a response form included wih the URGENT MEDICAL DEVICE CORRECTION letter and to notify any customers of this advisory. Hospira will also be supplementing the instructions for CLAVE use as an interim measure until a new method for bonding the CLAVE to the secondary port of the Plum cassettes can be devised and implemented.
- Initiated
- 2013-09-09
- Root cause
- Device Design
- Terminated
- 2016-12-06
- Distribution
- US (Nationwide) Distribution
- Recall number
- Z-2060-2016
- Recall status
- Terminated
- CDRH record ID
- 122414
- FDA record created
- 2016-06-29
- Recall event number