Z-2060-2018: Sonialvision Safire II, Model #: DAR-8000f Product Usage: This angiographic x-ray system device intende

Recall record Z-2060-2018 for Shimadzu Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2060-2018
Company in FDA record
Shimadzu Medical Systems
FDA record creation date
Jun 14, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When selecting serial radiography with a pulse rate of 7.5fps (including selecting a preset or changing the pulse rate using a temporary edit function) it was observed the indicated "irradiation time" on the X-ray Generator Console, as well as the "integral dose" on the external console of fluoroscopy, were reset to 0. There has been one report of this event.

Source facts

Action
Shimadf0u sent an Urgent Voluntary Medical Device Recall Notice letter dated January 2018. The letter identified the affected product, problem and actions to be taken. For questions contact Shimadzu representative or contact Shimadzu National Technical Support at 800-228-1429.
Initiated
2018-01-12
Root cause
Software design
Terminated
2022-02-24
Distribution
US Nationwide in the states of IL, WA, MI, TX, AZ, CA, LA, FL, SC, and NJ.
Recall number
Z-2060-2018
Recall status
Terminated
CDRH record ID
164342
FDA record created
2018-06-14
Recall event number
80067

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗