When selecting serial radiography with a pulse rate of 7.5fps (including selecting a preset or changing the
pulse rate using a temporary edit function) it was observed the indicated "irradiation time" on the X-ray
Generator Console, as well as the "integral dose" on the external console of fluoroscopy, were reset to 0. There has been one report of this event.
Source facts
Action
Shimadf0u sent an Urgent Voluntary Medical Device Recall Notice letter dated January 2018. The letter identified the affected product, problem and actions to be taken. For questions contact Shimadzu representative or contact Shimadzu National Technical Support at 800-228-1429.
Initiated
2018-01-12
Root cause
Software design
Terminated
2022-02-24
Distribution
US Nationwide in the states of IL, WA, MI, TX, AZ, CA, LA, FL, SC, and NJ.
Recall number
Z-2060-2018
Recall status
Terminated
CDRH record ID
164342
FDA record created
2018-06-14
Recall event number
80067