Two issues:
1) The MAC VU360 system may intermittently display an incorrect patient ID or visit number on the screen after scanning the patient s barcode. This may result in the ECG report being assigned to the incorrect patient.
2) The MAC VU360 may have the incorrect patient demographics appear on the patient banner. This may result in the ECG report being assigned to the incorrect patient.
Source facts
Action
The firm issued an Urgent Medical Device Correction letter on 06/28/2019 to tis consignees by mail. This letter informed the customer of the issues and provides instructions for continued use of the system. GE Healthcare will correct the system by upgrading the software.
Initiated
2019-06-28
Root cause
Software design
Terminated
2021-04-13
Distribution
Worldwide
Recall number
Z-2060-2019
Recall status
Terminated
CDRH record ID
174501
FDA record created
2019-07-23
Recall event number
83270