- FDA identifier
- Z-2060-2020
- Company in FDA record
- Bovie Medical Corporation
- FDA record creation date
- May 20, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Unexpected stress fractures on the shaft of the hand piece may result in fragmentation during clinical use.
Source facts
- Action
- On April 23, 2020 Apyx Medical (Formerly Bovie Medical) sent out an URGENT Medical Device Field Safety Notice identifying the problem, under what circumstances it can occur, the actions that should be taken by the customer/user in order to prevent risk to patients/users, and the actions planned by Apyx Medical to correct the problem. PRODUCTS: All lots of the Renuvion/J-Plasma Precise Hand piece with specified Catlog #'s that are not expired, and all lots of the Renuvion/J-Plasma Precise Open Handpiece with specified Catlog #'s that are not expired. ACTIONS TO BE TAKEN BY CUSTOMER/USER: Follow the cleaning instructions provided in the Instructions for Use (IFU) to remove coagulum/eschar from the blade during clinical use: For optimum performance, keep the distal end of the shaft free of debris. A damp gauze pad can be used for cleaning. Do not activate while cleaning the tip. Follow all specified cleaning instructions identified in the Field Safety Notification which are also in the product Instructions for Use (IFU's). Do Not modify or alter the tip of the device. Do Not use a cannula or a skin port that has tight interference with the outer diameter of the shaft of the device. Discontinue using the hand piece if during clinical use if a stress fracture develops and immediately report to Apyx Medical for return if any devices that exhibit this issue during clinical use. Customers should sign and return the customer response form via email indicating they have read and understand the instructions by May 1st of 2020.
- Initiated
- 2020-04-20
- Root cause
- Under Investigation by firm
- Terminated
- 2022-09-13
- Distribution
- Worldwide distribution - US nationwide including in the states of Puerto Rico, and the countries of CA, JO, CO, MX, BR, CR, PA, KW, CL, MA, ES, TH, PH, IQ, LB, DO, Japan, QA, AE, PE, AU, VN, IL, Lithuania, UAE, Belgium, Denmark, Austria, Poland, Germany, Slovakia, Spain, Cyprus, Switzerland, Turkey, UK, Italy, Bulgaria, Greece, Norway, Ireland, Hungary, and Croatia.
- Recall number
- Z-2060-2020
- Recall status
- Terminated
- CDRH record ID
- 181307
- FDA record created
- 2020-05-20
- Recall event number
- 85597