Open original FDA record ↗Recall record
- FDA identifier
- Z-2061-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Apr 25, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Product Failure-- When the 64-Channel System V5M is used in the 128-Channel mode, an over current fault occurs which shuts the system down.
Source facts
- Action
- Siemens sent a Customer Safety Advisory Notification letter on July 24, 2009. The letter identified the product, the problem, and the action to be taken by the customer. A software update will be conducted in the field. Until that update occurs, customers were instructed to image without the CW doppler to avoid a system shut down. Customers were instructed to contact their service support person with any questions. For questions regarding this recall call 650-694-5993.
- Initiated
- 2009-07-24
- Root cause
- Software design
- Terminated
- 2011-04-25
- Distribution
- Worldwide Distribution - USA including CA, CT, FL, MA, and TX and the countries of United Arab Emirates, Germany, Ecuador, Spain, France, Great Britian, Saudi Arabia, Sweden, Thailand, and South Africa.
- Recall number
- Z-2061-2011
- Recall status
- Terminated
- CDRH record ID
- 84555
- FDA classified
- 2011-04-25
- Recall event number
- 53053