Reason stated in the source

Over time, low-frequency vibrations can cause the pins within the female connectors on the Motor Controller to Data Acquisition Board ribbon cable to become partially displaced causing momentary high resistance that prevents or restricts data travelling through the cable during the high resistance event.

Source facts

Action
The firm, Philips, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated April 20, 2017 to its customers. The Customers letter inform them that Respironics California is recalling the Philips V60 Noninvasive Ventilators manufactured before 15 September 2015; that the recall is for all V60s with date of manufacture before 15 September 2015 and not for V60 Ventilators manufactured on or after 15 September 2015; of the recall reason, risk to health, how to identify devices subject to recall, and actions to be taken the customers. The customers were also instructed to complete and return the RECALL RETURN RESPONSE Acknowledgement and Receipt Form via email to: recall.response@philips.com or fax to: 1-877-499-7223 or mail to: Philips Healthtech, Mailstop# 4202, 3000 Minuteman Road, Andover, MA 01810-1032. On 5/9/17, Respironics has revised customer notification letter and began sending it out to their customers. The update includes the following actions to be taken: "appropriately trained personnel may continue using the V60" and "ensure an alternative form of ventilation device is available at all times, including intra-hospital transport (e.g. manual ventilation bag)." Philips will contact each consignee to schedule an appointment for this recall. Philip Engineers will remediate any affected V60 ventilators at the site. If you have any questions, please contact the Head of Q & R at 760-918-1067 or donald.sherratt@philips.com.
Initiated
2017-04-24
Root cause
Device Design
Terminated
2018-08-13
Distribution
Worldwide Distribution- US (nationwide) and countries of: Argentina, Australia, Austria, Bahrain, Barbados, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Hong Kong, Iceland, India, Indonesia, Iran (Islamic Republic Of), Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Korea, Republic Of, Kuwait, Lebanon, Libyan Arab Jamahiriya, Malaysia, Malta, Mexico, Monaco, Morocco, Namibia, Netherlands, Netherlands Antilles, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province Of China, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Viet Nam.
Recall number
Z-2061-2017
Recall status
Terminated
CDRH record ID
154944
FDA classified
2017-05-18
Recall event number
77084

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗