Z-2061-2018: nordicICE v 2.3.14 nordicICE is an image processing software package to be used by trained professionals

Recall record Z-2061-2018 for NordicNeuroLab AS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2061-2018
Company in FDA record
NordicNeuroLab AS
FDA record creation date
Jun 4, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The bug causes BOLD activation maps to be visualized as overlays without taking the coregistration into account. Any output created from these activation maps as overlays will not be adjusted according to the coregistration.

Source facts

Action
The firm notified their customers by email on July 3, 2015. Actions to be taken by the customer/User: We advise users to immediately stop using the BOLD module in nordiclCE and contact NordicNeuroLab for further instructions on how to upgrade the nordiclCE BOLD module, as described more fully below. For further questions, please call (262) 337-2909.
Initiated
2015-07-03
Root cause
Software design
Terminated
2020-09-16
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Australia , Belgium, Brazil, China, Czech Republic, Denmark, France, Germany, Greece, Hungary, India, Italy, Japan, Kuwait, Malaysia, Netherlands, Norway and Poland.
Recall number
Z-2061-2018
Recall status
Terminated
CDRH record ID
159628
FDA record created
2018-06-04
Recall event number
78467

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗