- Z-2062-2018
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA record creation date
- Jun 4, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Urea Nitrogen reagent kit demonstrated an increased incidence of Calibration Failures. Results cannot be generated by the system when calibration fails.
Source facts
- Action
- Date recall was initiated: 03-27-18 Customer communication was sent on 04-03-18 Urgent Medical Device Recall (CHC18-04.A.US) was distributed to all customers in the United States and an Urgent Field Safety Notice (CHC18-04.A.OUS) was distributed to all customers outside the United States who have received ADVIA Chemistry Urea Nitrogen reagent kit lots 408143 and 418290 (Lot 418290 was not sold in the US). These letters instruct the customers to discontinue use of reagent kit lots 408143 and 418290 and to contact their Customer Care Center for replacements kits. In the US the method of distribution is Fed Ex and outside the US the distribution is determined at the country level.
- Initiated
- 2018-03-27
- Root cause
- Material/Component Contamination
- Terminated
- 2021-02-03
- Distribution
- Nationwide and worldwide
- Recall number
- Z-2062-2018
- Recall status
- Terminated
- CDRH record ID
- 163963
- FDA record created
- 2018-06-04
- Recall event number
- 79932