Z-2062-2019: Merge PACS versions 7.3, 7.3.1, 8.0 and 8.0.1

Recall record Z-2062-2019 for Merge Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2062-2019
Company in FDA record
Merge Healthcare, Inc.
FDA record creation date
Jul 23, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for prior thumbnails to not display in reverse chronological order and images may not scroll in reverse chronological order when mammography stacked scrolling is enabled. Merge PACS provides image manipulation tools to enable users to view and compare images such as: linking, MPR, MIP, 3D image fusion/registration of CT, MR and PET; as well as CVR (Color Volume Rendering), measurements (linear distances, angles, areas, SUV, etc.), and annotations (for example, outline and label regions of interest, label spinal vertebrae).

Source facts

Action
Merge Healthcare notified customers on about 07/11/2019 via "Urgent Field Safety Notice" letter. The notice informed customers of two possible issues that may occur on Merge PACS and Merge OrthoPACS: 1. Prior thumbnails may not display in reverse chronological order. 2. Images may not scroll in reverse chronological order when mammography stacked scrolling is enabled. Use of an out of order image may result in the implementation of an incorrect or unnecessary treatment plan such as unnecessary medication, change in dosage of existing medication, or unnecessary medical procedures. It may also lead to progression of an existing patient condition, which may result in the progression of a disease state. Merge Healthcare will provide a software correction for the issue. A Merge Healthcare representative will contact each customer schedule the correction. In the interim, Merge Healthcare instructed customers that they can continue to use their Merge PACS/OrthoPACS devices by following the below instructions: 1. When pulling a series from thumbnail bar, confirm that the date of each series is visible in the viewport. The date will appear when overlays are configured to display the date. An administrator can assist in configuring Merge PACS/OrthoPACS if needed. 2. When utilizing the mammography scroll stacking tool, confirm that the date of each series is visible in the viewport. The date will appear when overlays are configured to display the date. An administrator can assist in configuring Merge PACS/OrthoPACS if needed. Customers were instructed to use the date, rather than the order the series are presented in, to determine chronological order. Additional instructions included to ensure all potential users of Merge PACS and Merge OrthoPACS have received this notice. Questions regarding the recall notice can be directed to Merge Customer Support at 877-741-5369.
Initiated
2019-07-11
Root cause
Software Design Change
Terminated
2023-03-17
Distribution
Nationwide distribution to Arizona, Arkansas, California, Colorado, Florida, Georgia, Hawaii, Illinois, Iowa, Kentucky, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Rhode Island, South Carolina, Texas, Utah, Virginia, Wisconsin. International distribution to Belgium and Canada.
Recall number
Z-2062-2019
Recall status
Terminated
CDRH record ID
174646
FDA record created
2019-07-23
Recall event number
83300

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗