Z-2062-2020: Oryx Cervical Screw Caddy, Model Number BTA-172, as part of the Oryx Cervical Plate System - Product Usag

Recall record Z-2062-2020 for Innovasis, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2062-2020
Company in FDA record
Innovasis, Inc
FDA record creation date
May 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cervical screw caddies have incorrect text markings for the 4.5mm Fixed and Variable rescue screws. The caddy location labeled as "Fixed" should be labeled as "Variable" and the caddy location labeled as "Variable" should be labeled as "Fixed. If user installs a fixed rescue screw instead of variable, the screw head may not be flush, it could back out, which could interfere with the cover screw.

Source facts

Action
On 04/07/202, the firm emailed the "URGENT: ADVISORY NOTICE", to consignees informing them of their plan to correct this issue with a three-step approach: 1) Make the consignee aware of the issue. Further, consignees were asked to notify any necessary case reps and/or surgeon users who may need to understand this distinction in the next week, before the firm issues a temporary fix. 2) The firm is initiating the following "temporary fix": Future orders will include reworked caddies with the words "FIXED" and "VARIABLE" removed. By doing this, the default organization is by color and by laser marking on the parts themselves. Fixed screws can be identified by the color (gold) or the first letters of the Part No. on the screw head (APF). Variable screws can be identified by the green color or by Part No. starting with APV. If/when you receive one of these temporary caddies, please pull the screws out of your current caddies and place them in one of the temporary caddies and return the incorrect caddy. Our Customer Service department will provide return shipping labels as necessary. 3) Another recall letter will follow requesting that consignees swap screws from temporary caddies into the correct caddies, and return temporary caddies. On 04/13/2020, the firm emailed "'URGENT: MEDICAL DEVICE RECALL" notices and Response Forms to consignees. The firm is issuing a "temporary Fix": Future orders will include reworked caddies with the words "FIXED" and "VARIABLE" corrected by way of laser engraving. These caddies will be in a conforming state. Upon receiving one of these reworked caddies, please pull the screws out of your current (non-conforming) caddies and place them in one of the reworked (conforming) caddies and return the original non-conforming caddy. The firm will be ordering new caddies that will then be swapped out with the reworked caddies. The firm requests consignees complete and return Product Removal Forms. Product Removal Forms will indicate ho
Initiated
2020-04-07
Root cause
Process control
Terminated
2022-01-26
Distribution
US Nationwide distribution including in the states of CA, AL, LA.
Recall number
Z-2062-2020
Recall status
Terminated
CDRH record ID
181263
FDA record created
2020-05-20
Recall event number
85575

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗